Shunt Valve Settings

Quick visual reference for common programmable shunt valves. Click any image for full detail.

Aesculap programmable shunt valve settings reference
Aesculap PSV (proGAV / proSA)
Orient using radiolucent margin / dashed lines; star/marker points to setting.
AesculapproGAVproSA
Orientation: proGAV = flat radiolucent margin → setting (example: 3 cm H2O). proSA = dashed lines; star → setting (example: 30 cm H2O). proSA 2 = dashed line; triangle markers; star → number (example: 14 cm H2O).
Codman Certas shunt valve setting reference
Codman Certas
Arrow + circle for orientation; rectangle/ball points to setting.
CodmanCertas
Orient: arrow + circle. Read: rectangle with ball → setting (example: 4).
Codman Hakim shunt valve setting reference
Codman Hakim
Circle/line + circle for orientation; radiolucent notch points to setting.
CodmanHakim
Orient: circle/line + circle. Read: radiolucent notch → setting (example: 30).
Medtronic CSF shunt valve setting reference
Medtronic CSF
Fixed valve with 4 fixed settings.
MedtronicCSF
Fixed valve: 4 fixed settings.
Medtronic Strata II shunt valve setting reference
Medtronic Strata II
Arrow + circles for orientation; notch points to setting.
MedtronicStrata II
Orient: arrow + circles. Read: notch → setting (example: 1.5).
Medtronic Strata II Small shunt valve setting reference
Medtronic Strata II Small
Arrowhead + circle for orientation; notch points to setting.
MedtronicStrata IISmall
Orient: arrowhead + circle. Read: notch → setting (example: 0.5).
Medtronic Strata NSC shunt valve setting reference
Medtronic Strata NSC
Triangle + circle for orientation; notch points to setting.
MedtronicStrataNSC
Orient: triangle + circle. Read: notch → setting (example: 0.5).
Medtronic Strata MR shunt valve setting reference
Medtronic Strata MR
Arrowhead + ellipse for orientation; notch points to setting.
MedtronicStrataMR
Recall: Currently not used due to FDA Class I recall.
Sophysa Polaris and SM8 shunt valve setting reference
Sophysa (Polaris / SM8)
Polaris (left) and SM8 (right) references with distinct orientation markers.
SophysaPolarisSM8
Polaris (left): curvilinear line + rectangle; square → 1 of 5 (example: 3).